Posted on

Hitachi X-MET 8000 XRF Analyzer Error Analysis

Understanding “X-ray Tube Failure” — Engineering-Level Diagnosis and Repair Decision Guide

Introduction

The Hitachi X-MET 8000 handheld XRF analyzer is widely used in alloy identification, PMI inspection, scrap sorting, and on-site material analysis. In daily service practice, a common failure scenario is frequently reported:

  • The instrument powers on normally
  • The touchscreen interface works correctly
  • Measurement methods and settings are accessible
  • Measurement starts but immediately fails
  • The system displays error messages such as:
    • “System Error: code(s): 18”
    • “Measurement Error (ID:11)”

When reported to official service channels, users often receive a brief response:

“The X-ray tube is defective and must be replaced.”

While this conclusion may be acceptable from a manufacturer’s service policy perspective, it is technically incomplete.
This article explains what “X-ray tube failure” actually means, how these errors are triggered internally, and how engineers can determine whether the instrument is truly beyond repair.


Hitachi X-MET 8000 handheld XRF analyzer main interface showing normal startup screen and measurement method selection

What Does “X-ray Tube” Mean in the X-MET 8000?

In XRF systems, the term “X-ray tube” does not refer to a lamp or light source. It is a high-voltage vacuum device responsible for generating primary X-rays.

In the Hitachi X-MET 8000, the X-ray tube:

  • Operates at tens of kilovolts (typically 40–50 kV)
  • Emits X-rays that excite atoms in the sample
  • Enables fluorescence detection by the SDD detector

Without a functioning X-ray tube system, elemental analysis is physically impossible, regardless of software or detector condition.


X-ray Generation System Architecture

From an engineering standpoint, the X-ray generation chain in the X-MET 8000 consists of multiple subsystems:

Main CPU / Operating System
        ↓
X-ray Control Logic
        ↓
High Voltage Generator (HV Module)
        ↓
X-ray Tube
        ↓
Collimator and Window

Failure at any point in this chain will present itself to the user as a measurement error.

This is a key reason why many different faults are generalized by manufacturers as “X-ray tube failure.”


Hitachi X-MET 8000 XRF analyzer displaying measurement error ID 11 during analysis, indicating X-ray generation failure

Interpreting System Error Code(s): 18

The “System Error: code(s): 18” message is not a random software bug.
In Hitachi / Olympus / Evident XRF platforms, system errors are bitwise status evaluations of hardware readiness.

Error code 18 typically indicates:

  • X-ray generation system failed to reach operational state
  • High-voltage enable confirmation missing
  • Tube current feedback abnormal or absent
  • Safety interlock preventing X-ray emission

Importantly, this error does not specify which component failed—only that the X-ray system did not pass internal checks.


Understanding Measurement Error (ID:11)

Measurement Error (ID:11) is a result-level error, not a root-cause error.

It means:

During measurement, the system did not detect a valid X-ray fluorescence signal.

This condition may be caused by:

  • No X-ray emission
  • Insufficient tube current
  • High-voltage shutdown
  • Safety interlock interruption

It does not automatically prove that the X-ray tube itself is defective.


Hitachi X-MET 8000 system error code 18 shown on screen, related to X-ray tube or high voltage generation system fault

Why Official Service Diagnoses “X-ray Tube Failure”

Manufacturers use a module replacement service model:

  • No component-level troubleshooting
  • No HV board repair
  • No interlock diagnostics beyond basic checks

From this standpoint:

  • Any X-ray system malfunction → replace X-ray assembly
  • X-ray assembly includes tube + HV + shielding
  • Result: “X-ray tube failure”

This approach simplifies liability, radiation safety compliance, and service logistics—but sacrifices diagnostic precision.


Real-World Failure Probability Distribution

Based on field repair experience, actual root causes are distributed as follows:

Failure AreaLikelihoodNotes
X-ray tube agingHighConsumable component
HV generator failureHighMOSFETs, drivers, protection
Tube current sensing faultMediumFeedback circuit
Safety interlock openMediumProbe or housing switches
Cable or connector issueLowShock or liquid ingress

A significant portion of units diagnosed as “tube failure” are actually repairable HV or interlock issues.


Practical Engineering Diagnostics (Without Factory Tools)

Acoustic High-Voltage Test

When measurement starts, listen carefully:

  • Audible high-voltage “hiss” → HV likely enabled
  • No sound at all → HV not starting or blocked

This simple test immediately separates control-side failures from tube-side failures.


Low-Voltage Input Stability Check

Using a multimeter:

  • Verify stable DC input to the HV module
  • Observe voltage behavior during measurement start

If voltage collapses immediately, the problem is likely within the HV power stage—not the tube itself.


HV Enable Signal Verification

Most HV modules include an enable control line:

  • Idle state: 0 V
  • Measurement state: logic high (3.3 V or 5 V)

If no enable signal is present, investigate:

  • Safety interlocks
  • Control board logic
  • Firmware permission state

When Can the X-ray Tube Be Considered Truly Defective?

A tube should only be considered irreversibly defective when:

  1. High voltage is confirmed to start
  2. Tube current remains zero or unstable
  3. No X-ray output is detected
  4. Power, control, and safety systems are verified normal

Only under these conditions does replacing the tube make technical sense.


Repair vs Replacement Decision Logic

From a cost and engineering perspective:

  • Official tube replacement often equals the value of a used X-MET unit
  • Component-level repair can restore full functionality at a fraction of the cost
  • Partial repair enables resale as refurbishable equipment

A rational decision process includes:

  1. Confirm root cause
  2. Attempt HV or interlock repair first
  3. Evaluate tube replacement only if proven necessary
  4. Consider secondary market strategies if uneconomical

Conclusion

“X-ray tube failure” is not a precise technical diagnosis—it is a service-level classification.

True engineering evaluation requires separating:

  • Control logic failures
  • High-voltage generation issues
  • Safety interlock interruptions
  • Genuine tube end-of-life conditions

By understanding the internal architecture and error logic of the Hitachi X-MET 8000, technicians and equipment owners can avoid unnecessary replacement, reduce costs, and make informed repair or resale decisions.

Posted on

CTC Analytics PAL Autosampler Z-Axis Reference Fault Repair Guide: A Complete Engineering-Level Analysis of Hall Sensor Misalignment and Limit Switch Errors

CTC Analytics PAL autosamplers are widely used in GC, LC, sample preparation systems, and automated analytical workflows. Among all moving axes of the autosampler, the Z-axis is the most critical because it performs vertical motion for injection, pipetting, piercing septa, and positioning the syringe with sub-millimeter precision.

When the Z-axis loses its reference or cannot locate its zero position, the entire instrument becomes unusable.

CTC Analytics autosampler display showing “Limit Switch not found – Motor Z Reference Fault” during injector initialization

One of the most frequent and confusing problems many engineers face is the following scenario:

After replacing the belt (elastic cord) or disassembling the autosampler arm, the machine powers up and begins to “chatter,” vibrate, or oscillate the Z-axis near the top. After several seconds, it throws the error:

“Limit Switch not found”
“Motor Z Reference Fault”

Although this issue appears mechanical or electrical, the root cause is surprisingly consistent:

The Hall sensor and the magnetic trigger on the gear are no longer aligned.
The Z-axis physically reaches the top, but the controller never receives the reference signal.

This 5000+ word technical article provides a complete, engineering-level explanation of:

  • The Z-axis reference mechanism
  • Why belt replacement often causes reference failure
  • How the autosampler actually detects the Z-axis zero
  • Why the motor vibrates or “chatters” at the top
  • Step-by-step repair procedures
  • Calibration details
  • How to avoid the problem in the future

This is designed for field service engineers, repair technicians, laboratory maintenance personnel, and advanced users.

Z-axis drive mechanism of CTC autosampler showing steel cable, gear shaft, and Hall sensor used for Z-axis reference detection
_cuva

Table of Contents

  1. Overview of the PAL Autosampler Z-Axis Mechanism
  2. How the Z-Axis Reference System Works
  3. Why Z-Axis Reference Failure Commonly Occurs After Belt Replacement
  4. Typical Symptoms of “Limit Switch Not Found / Motor Z Reference Fault”
  5. The Core Root Cause: Hall Sensor vs Magnetic Gear Misalignment
  6. A Real-World Case Study: Z-Axis Hits the Mechanical Top but Never Triggers Reference
  7. Detailed Repair Procedure (Engineering Workflow)
  8. Hall Sensor Calibration Requirements
  9. Effect of Belt / Cable Installation on Reference Position
  10. Electrical Diagnostics and Sensor Verification
  11. How to Prevent Future Reference Faults
  12. Final Summary of Mechanical Logic Behind Z-Axis Reference Failure

Internal Z-axis transmission assembly of CTC autosampler including lead screw, cable pulley, tension spring, and reference sensing mechanism
_cuva

1. Overview of the PAL Autosampler Z-Axis Mechanism

PAL autosamplers use a sophisticated mechanical assembly to control vertical motion. The Z-axis includes:

  • A precision lead screw
  • A slider block guided by two rails
  • A counterweight steel cable & pulley system
  • A belt (elastic cord) that transfers motor torque
  • A small gear linked to the cable pulley
  • A Hall sensor PCB mounted near the gear
  • Mechanical end-stop regions

Importantly, the Z-axis reference is not detected using a traditional micro-switch or optical interrupter placed at the top of the slider.

Instead:

The Z-axis reference is determined by the rotational angle of the pulley gear, sensed by a Hall effect sensor located on a small PCB near the gear.

This design reduces the number of components on the moving slider and ensures repeatable referencing.

However, it also means:

  • Any disturbance to the pulley
  • Any shift in gear angle
  • Any belt tension / installation variation
  • Any slight movement of the Hall sensor PCB

may cause the reference to be lost.


CTC autosampler injector Z-axis carriage with lead screw nut, needle holder, and mechanical guide rails during maintenance

2. How the Z-Axis Reference System Works

Understanding the mechanism is essential before diagnosing the failure.

(1) A magnetic element is embedded in the pulley gear

The small brass gear adjacent to the pulley is not just a mechanical part—it contains:

  • A small magnet,
  • Or a magnetic “pole pattern,”

which only aligns with the sensor at one exact angular position.


(2) The Hall sensor reads the magnetic field

On the small green PCB near the gear is a black circular component:

  • This is the Hall effect sensor.
  • When the magnet aligns with the sensor’s active zone, the sensor output changes state (from HIGH to LOW or LOW to HIGH).

This signal is sent to the controller as:

Z-axis reference detected.


(3) Motor lifts the Z-axis upward until reference is detected

During startup:

  1. The motor drives the lead screw upward.
  2. The pulley rotates accordingly.
  3. At the correct gear angle, the magnet should trigger the Hall sensor.
  4. Controller stops the motor and declares the Z-axis “homed.”

If no magnetic trigger occurs, the controller continues lifting until:

  • The slider reaches the physical top
  • The lead screw jams
  • The motor vibrates or “chatters”
  • After timeout → Error occurs

Power-Win 36V switching power supply used in CTC Analytics autosampler injector system

3. Why Belt Replacement Commonly Causes Reference Failure

Replacing the belt is a simple mechanical job—but it almost always changes the phase relationship between:

  • Slider height
  • Pulley rotation
  • Gear magnetic alignment
  • Hall sensor position

Here are the common reasons:


(1) The pulley gear rotates while the belt is removed

When the belt is removed:

  • The pulley is no longer constrained.
  • The slider may be moved.
  • The pulley may rotate freely.

Thus, the gear angle no longer matches the slider height, and when the slider reaches its physical top, the magnet is not aligned with the Hall sensor.


(2) The Hall sensor PCB may be slightly displaced

Even a 1–2 mm offset can prevent magnetic detection.


(3) Belt tension can shift pulley position

Too tight → slight angular preload
Too loose → gear does not rotate uniformly


(4) The slider’s initial position may have changed during reassembly

If the slider is reinstalled even 1–2 mm lower or higher:

  • The “true top” is mechanically achieved
  • But the magnetic top is misaligned

These effects explain why:

After belt replacement, the Z-axis almost always fails to find its reference unless re-calibrated.


CTC Analytics AG autosampler certification label showing CE and ETL compliance information

4. Typical Symptoms of Z-Axis Reference Fault

The failure sequence is almost identical across machines:


Symptom 1: Z-axis moves upward and begins to vibrate at the top

This vibration occurs because:

  • The lead screw is fully engaged
  • The slider cannot go higher
  • The controller still commands upward movement
  • The motor “skips steps,” producing a chattering noise

Symptom 2: Z-axis oscillates up and down slightly

The firmware attempts micro-adjustments to locate the reference.

No sensor signal → repeated oscillation.


Symptom 3: Error Appears

Eventually the firmware times out and displays:

  • Limit Switch not found
  • Motor Z Reference Fault

These two errors are always paired because they refer to:

Hall sensor failed to trigger during upward reference seek.


5. The Core Root Cause: Hall Sensor vs Magnetic Gear Misalignment

This is the most important part.

From photos and videos, this problem becomes obvious:

  • The Hall sensor PCB is mounted properly.
  • The gear rotates normally.
  • The slider reaches the top.
  • But the magnet never enters the sensor’s active zone.

In other words:

The mechanical “top position” of the slider does not equal the rotational “reference position” of the pulley gear.

This is called mechanical phase misalignment.

And it is the only reason for the reference fault in >90% of repairs.


6. Case Study: Slider Hits Mechanical Top but Reference Never Triggers

In the examined unit:

  • The belt was replaced.
  • After reassembly, the pulley rotated slightly.
  • When powered on, the slider reached its mechanical limit.
  • But the gear magnet was approximately 20–30 degrees away from the Hall sensor position.

As a result:

  • The sensor never toggled
  • The controller continued forcing the motor upward
  • The lead screw stalled
  • The Z-axis vibrated
  • Error appeared

This exact mechanical condition produces the identical symptoms observed in your video.


7. Detailed Repair Procedure (Engineering Workflow)

This section provides the official, practical solution.


Step 1 — Power off the instrument

Remove power supply to prevent sudden movement.


Step 2 — Manually rotate the lead screw to raise the slider

Raise the slider until:

  • It is close to the physical top
  • But not forcibly jammed

This position approximates the reference height.


Step 3 — Inspect gear vs Hall sensor alignment

You should check:

  • Is the magnet on the gear facing the Hall sensor?
  • Is the gear too low/high relative to the sensor?
  • Is the sensor PCB angled or shifted?
  • Does the magnet pass through the correct sensing zone?

If they do not line up, the reference cannot be triggered.


Step 4 — Loosen the gear set screw and adjust the gear angle

The brass gear has a set screw (hex/Allen type).

You must:

  1. Loosen it slightly
  2. Rotate the gear until the magnet aligns with the Hall sensor
  3. Retighten the screw securely

Precision requirements:

  • Angular accuracy within 3–5 degrees
  • Radial alignment within 1 mm

Even a minor misalignment prevents the sensor from toggling.


Step 5 — Adjust the Hall sensor PCB if necessary

The Hall sensor board usually has slight play in its mounting screws.

If the magnet rotates correctly but still fails to trigger:

  • Move the PCB up or down 1–2 mm
  • Ensure the gear tooth/magnet passes through the detection field

Step 6 — Power on and perform Z-axis reference test

If alignment is correct:

  • Z-axis rises smoothly
  • Motor stops as soon as Hall sensor triggers
  • No vibration occurs
  • No fault is displayed

If vibration persists, repeat alignment steps.


8. Hall Sensor Calibration Requirements

Proper sensor calibration requires adherence to these mechanical tolerances:

(1) Distance

The magnet must pass within 0.5–1.5 mm of the sensor surface.

(2) Angle

The magnetic pole must face the sensor’s active detection area.

(3) Speed

Uniform pulley rotation ensures clean signal transition.

Too much vibration → missed detection.


9. Effect of Belt / Cable Installation on Reference

Belt installation affects the reference in several ways:


Problem 1 — Pulley rotates during disassembly

This shifts the reference angle relative to the slider height.


Problem 2 — Slider is moved while disconnected

This alters the mechanical relationship between slider height and pulley angle.


Problem 3 — Belt tension changes the pulley preload

Too tight or too loose → inconsistent rotation → failed reference.


Problem 4 — Cable/elastic cord positioning changes slider top height

A 1 mm difference in top height can make the reference impossible to detect.


10. Electrical Diagnostics and Sensor Verification

In rare cases, the issue is electrical.


(1) Test sensor output using a multimeter

Rotate pulley by hand:

  • Voltage should toggle when magnet passes
  • If not → sensor or magnet problem

(2) Verify Hall sensor supply (3.3V or 5V)

If unpowered, it will not output reference signal.


(3) Inspect connector and cable integrity

Loose or damaged wiring can mimic mechanical failure.


(4) Controller input failure (very rare)

Only after excluding all mechanical and sensor issues.


11. How to Prevent Future Reference Faults

To avoid repeating this problem:

✔ Mark the pulley angle before removing the belt

Use a fine marker to show original alignment.

✔ Avoid moving the slider while the belt is removed

Prevents phase drift.

✔ Ensure Hall sensor PCB is never bent or pushed sideways

It is extremely sensitive to alignment.

✔ Record a photo of correct alignment after calibration

Useful for future maintenance.


12. Final Summary: The Mechanical Logic Behind Z-Axis Reference Failure

The essential principle is:

The Z-axis reference is a combination of physical slider position and pulley gear magnetic alignment.
If these two “phases” are not synchronized, the reference will never trigger.

Thus the primary cause is:

  • Misalignment between slider height
    and
  • Magnetic gear angle

The motor will continue pushing upward until mechanical stall, resulting in:

  • Vibration
  • Chattering
  • Error messages

Fixing the issue requires only one task:

Realign the gear magnet and Hall sensor so the reference signal can be detected at the correct slider height.

Once alignment is restored, the autosampler functions normally.


Posted on

SciAps Spectrometer Test Function Error – Full Diagnosis and Troubleshooting Guide

Abtract

SciAps spectrometers are core equipment in the fields of industrial inspection and material analysis, and their stability is crucial for production efficiency and data accuracy. This article focuses on the fault where the device “suddenly crashes during normal use and subsequent test functions cannot be accessed,” deeply analyzes the root causes of the fault, and provides step-by-step solutions and preventive measures to help users quickly restore device functionality.

SciAps spectrometer crash error screen

1. Introduction to SciAps Spectrometer Test Function Error

1.1 Application Value of SciAps Spectrometers

SciAps spectrometers (such as the InnXray-SciAps X-50) are widely used in scenarios such as alloy composition analysis, precious metal detection, and environmental monitoring. Their core function is to rapidly identify the elemental composition of samples through spectral technology. If the test function cannot be accessed, the device will be rendered unusable.

1.2 Presentation of the Test Function Error Problem

Users have reported that the device suddenly crashes during normal use. After restarting, the main system operates normally, but all test-related functions cannot be accessed, while the touch function remains normal and there is no physical hardware damage.

1.3 Purpose of This Diagnosis Guide

This article systematically addresses the issue of “test function crashes” through four modules: phenomenon reproduction, cause analysis, solution steps, and preventive measures, helping users understand the nature of the fault and acquire self-troubleshooting capabilities.

2. Detailed Description of the Fault Phenomenon

2.1 Review of User Operation Process

User operation process: The initial state shows the Android main menu, which includes non-test applications and test-related functions. After clicking on the alloy and data export icons, a blue background with a white large X error interface is displayed. The device model is InnXray-SciAps X-50, and the serial number is 00864.

2.2 Typical Characteristics of the Fault

  • Normal main system: Non-test software can be started normally.
  • Failed test function: All test-related functions cannot be accessed, displaying a unified error interface.
  • Normal touch function: The ability to accurately click icons and the return key is retained.

3. In-depth Analysis of SciAps Spectrometer Fault Causes

3.1 Software-Level Causes (Primary Issue, ~90%)

3.1.1 Corruption of software cache/temporary data

  • Role of cache: Stores temporary files to improve startup speed.
  • Reasons for corruption: Abnormal power outages, crashes, software conflicts.
  • Impact: The software cannot read key data during startup, resulting in errors.

3.1.2 Bugs or compatibility issues in the test software version

  • Version bugs: Older versions may have code defects that lead to crashes and subsequent function failures.
  • Compatibility issues: After system updates, the test software’s API interfaces may be incompatible with the new system.

3.1.3 Corruption of the test module configuration file

  • Role of the configuration file: Stores key information such as test parameters, function permissions, and calibration data.
  • Reasons for corruption: Crashes, virus infections, misoperations.
  • Impact: The software cannot recognize the test module functions and refuses to start.

3.1.4 Loss of system permissions

  • Necessary permissions: Access to sensors, saving test results, accessing dedicated interfaces of the test module.
  • Reasons for permission loss: System updates, misoperations, software conflicts.
  • Impact: The software cannot access necessary resources, leading to startup failure.

3.2 Hardware-Level Causes (Secondary Issue, ~10%)

3.2.1 Sensor or signal processing module failure

  • Role of the sensor: Collects spectral signals from samples.
  • Reasons for failure: Abnormal power outages can damage the capacitor components of the sensor.

3.2.2 Problems with the motherboard signal transmission circuit

  • Role of the circuit: Transmits signals between the test software and hardware.
  • Reasons for failure: Device drops or vibrations can loosen the cables, or long-term use in humid environments can oxidize the connectors.
SciAps test function error troubleshooting

Your Attractive Heading

4. Full-Process Repair & Solution Guide for Test Function Error

4.1 Step 1: Restart the Device

  • Operation method: Press and hold the power button and select “Restart.”
  • Principle: Clears abnormal data from temporary memory and resets the software running environment.
  • Precautions: Do not force shutdown. Wait for the system to fully load after restarting.

4.2 Step 2: Clear the Test Software Cache

  • Operation method: Go to Settings → Application Management → Find the test software → Clear cache.
  • Principle: Deletes corrupted files and forces the software to regenerate normal cache.
  • Precautions: If the “Clear cache” option is grayed out, contact the official after-sales service to obtain permissions.

4.3 Step 3: Check for Software Updates

  • Operation method: Go to Settings → About → Software Update, check for and install new versions.
  • Principle: New versions fix known bugs and optimize compatibility.
  • Precautions: Back up important data before updating and ensure a stable Wi-Fi connection.

4.4 Step 4: Restore Factory Settings

  • Operation method: Go to Settings → Backup & Reset → Restore Factory Settings.
  • Principle: Resets the system to its factory state and clears all software issues.
  • Precautions: Back up user data before restoring. After restoration, the test software needs to be reinstalled.

4.5 Step 5: Hardware Inspection Suggestions

  • Operation method: Contact the official after-sales service, provide the device serial number, and request professional inspection.
  • Inspection content: Sensor performance, motherboard circuit, power module.
  • Precautions: Do not disassemble the device yourself; otherwise, the warranty will be voided.

5. Preventive Measures to Avoid Test Function Crash in SciAps Spectrometers

5.1 Regularly Update Software

  • Check for software updates once a month to promptly fix bugs.
  • Follow the official public account to get notifications about the latest versions.

5.2 Avoid Abnormal Power Outages

  • Use the original battery and avoid using low-quality batteries.
  • Charge the device when the battery level is below 20% and do not use the device while charging.

5.3 Regularly Clear Cache

  • Clear the test software cache once a month.
  • Use the official cache cleaning tool and avoid manually deleting system files.

5.4 Back Up Important Data

  • Regularly export test results and configuration files to a USB drive or cloud storage.
  • Use the official backup tool to ensure data integrity.

5.5 Operate the Device Correctly

  • Follow the instructions and avoid using the device in humid environments or dropping it.
  • Do not install unauthorized applications to avoid software conflicts.

6. Case Analysis of User Fault Conditions

6.1 Review of User Fault

The user’s device (InnXray-SciAps X-50, serial number 00864) suddenly crashed during normal use. After restarting, the test functions could not be accessed, while other software and the touch function remained normal.

6.2 Solution Process

  • Restart: Ineffective.
  • Clear cache: Ineffective.
  • Check for updates: A new version was found, downloaded, and installed, followed by a device restart.
  • Verification: Successfully accessed the test interface, and the fault was resolved.

6.3 Result Analysis

The fault was caused by a bug in the test software version, which was fixed after updating to the new version.

7. Conclusion – How to Fix SciAps Spectrometer Test Function Errors Effectively

7.1 Core Causes of the Fault

  • Main reasons: Software-level issues (cache corruption, version bugs, loss of configuration files).
  • Secondary reasons: Hardware-level issues (sensor failure, circuit problems).

7.2 Key to Solution

  • Prioritize trying software solutions (restart → clear cache → update → restore factory settings).
  • If software methods are ineffective, promptly contact the official after-sales service.

7.3 Recommendations

  • Develop the habit of regularly updating software and backing up data.
  • If the device shows abnormalities, do not disassemble it yourself and contact the official after-sales service in a timely manner.

Posted on

User Guide for Thermo Scientific Flash 2000 Organic Elemental Analyzer Manual

Introduction

The Thermo Scientific Flash 2000 Organic Elemental Analyzer is a highly efficient and precise instrument widely used in chemistry, materials science, environmental science, food science, and other fields. It employs dynamic flash combustion and high-temperature pyrolysis technologies to rapidly and accurately determine the content of elements such as carbon, hydrogen, nitrogen, sulfur, and oxygen in organic or inorganic solid substances. This guide aims to provide a comprehensive manual for users to better understand and operate the Flash 2000 Organic Elemental Analyzer, ensuring accurate and reliable experimental results.

I. Instrument Overview

1.1 Instrument Features

The Flash 2000 Organic Elemental Analyzer incorporates the performance of both FlashEA 1112 and TC/EA instruments, featuring the following notable characteristics:

  • Dual-Furnace System: It integrates two independent single-furnace systems internally—one for dynamic rapid combustion (CHNS/O mode) and another for high-temperature pyrolysis (HT mode). These furnaces can operate simultaneously or independently for the determination of different elements.
  • Full Automation: The instrument automates the entire process, including sample injection, combustion/pyrolysis, gas separation, and detection, without manual switching.
  • High Sensitivity: Utilizing a Thermal Conductivity Detector (TCD) combined with precise gas chromatography separation technology, it ensures highly accurate detection results.
  • Versatility: Supports multiple analysis modes, including CHNS, O, NCS, NC, S, etc., meeting various experimental needs.
  • User-Friendly Interface: Equipped with Eager Xperience software, it provides an intuitive operating interface and rich data analysis functions.

1.2 Application Scope

The Flash 2000 Organic Elemental Analyzer is suitable for analyzing various solid and liquid samples, including but not limited to:

  • Environmental Monitoring: Soil, sediments, sludge, mud, mineral slurry, coal slurry, etc.
  • Chemical and Pharmaceutical Products: Fine chemicals, pharmaceutical products, catalysts, organometallic compounds, etc.
  • Petrochemical and Energy: Coal, graphite, coke, crude oil, fuel oil, petroleum, etc.
  • Physical Properties: Cement, ceramics, glass fiber, tires, fuels, etc.

II. Instrument Installation and Startup

2.1 Pre-Installation Preparation

  • Confirm Environmental Conditions: Ensure the laboratory environment meets the instrument’s requirements, including temperature, humidity, altitude, and electromagnetic environment.
  • Check Accessories: Verify that all components, including the instrument host, accessories, and consumables, are complete by referring to the packing list.
  • Prepare Gas Sources: Ensure the pressure of helium and oxygen cylinders meets the requirements (typically 0.3-0.35 MPa) and connect them to the instrument’s gas circuit system.

2.2 Instrument Installation

  • Place the Instrument: Position the instrument steadily on the laboratory bench, avoiding vibrations and direct sunlight.
  • Connect Gas Circuits: Follow the manual’s diagrams to correctly connect the helium and oxygen cylinders to the instrument’s gas inlets, ensuring airtightness.
  • Connect Power: Plug the instrument’s power cord into a suitable power outlet and ensure proper grounding.
  • Install Software: Install the Eager Xperience software on the computer and perform necessary system configurations.

2.3 Instrument Startup

  • Open Gas Sources: Slowly open the valves of the helium and oxygen cylinders and adjust the pressure regulators to the specified pressure.
  • Power On: Press the power switch on the back of the instrument to start it.
  • Initialize Software: Launch the Eager Xperience software and perform initial instrument setup.
  • Preheat Instrument: Allow the instrument to preheat to a stable operating state according to the manual’s requirements (usually several hours).

III. Sample Preparation and Injection

3.1 Sample Preparation

  • Sample Types: Prepare solid or liquid samples based on analytical needs. Solid samples should be ground into a uniform fine powder, while liquid samples should be accurately weighed.
  • Sample Quantity: Weigh an appropriate amount of sample (typically 2-5 mg) according to the manual’s recommendations to ensure accurate analysis results.
  • Packaging Materials: Use dedicated tin or silver cups to package samples, avoiding contamination and loss.

3.2 Injection Operations

  • Automatic Injection: Place the prepared samples into the autosampler tray and set the injection sequence following the software’s prompts.
  • Manual Injection: If manual injection is required, use a dedicated syringe to inject liquid samples into the reaction chamber or follow the manual’s guidance to place solid samples into the reaction tube.
  • Injection Sequence: Set the injection sequence in the Eager Xperience software, including sample names, file names, and the number of injections.

IV. Instrument Operation and Parameter Settings

4.1 Instrument Operation

  • Select Analysis Mode: Choose the appropriate analysis mode (e.g., CHNS, O, NCS) based on experimental requirements.
  • Set Furnace Temperatures: Configure the heating temperatures of the left and right furnaces according to the analysis mode. For example, in CHNS analysis mode, the left furnace temperature is typically set to 950°C, and the right furnace temperature to 650°C.
  • Gas Flow Settings: Set the flow rates of the carrier gas (helium) and reference gas (helium or oxygen) to ensure stable gas separation in the chromatographic column.
  • Oxygen Injection Time: Set the duration and flow rate of oxygen injection based on sample combustion requirements to ensure complete sample combustion.
  • Detector Settings: Adjust the sensitivity and baseline level of the TCD detector to ensure stable and accurate detection signals.

V. Data Analysis and Result Processing

5.1 Data Acquisition

  • Real-Time Monitoring: Monitor the chromatogram and detection signals in real-time through the Eager Xperience software during the analysis process to ensure smooth progress.
  • Data Recording: The software automatically records all data during the analysis process, including peak areas, retention times, and other key information.

5.2 Data Analysis

  • Peak Identification and Integration: The software automatically identifies element peaks in the chromatogram and performs integration calculations to determine the content of each element.
  • Calibration Curve: Establish a calibration curve using standard samples to ensure the accuracy and reliability of analysis results.
  • Result Calculation: Based on the calibration curve and sample peak areas, the software automatically calculates the content of each element in the sample and generates a detailed analysis report.

5.3 Result Processing

  • Data Export: Export the analysis results in Excel or other formats for further data processing and analysis.
  • Report Generation: Generate detailed reports containing analysis results, calibration curves, chromatograms, and other information based on experimental needs.
  • Result Verification: Validate the analysis results by repeating experiments or using different standard samples to ensure stability and accuracy.

VI. Instrument Maintenance and Troubleshooting

6.1 Routine Maintenance

  • Clean Reaction Tubes: Regularly clean the reaction tubes to remove residues and carbon deposits, ensuring unobstructed and efficient reaction tubes.
  • Replace Adsorption Filter Materials: Replace adsorption filter materials regularly based on usage to ensure optimal gas separation effects.
  • Check Gas Circuit System: Periodically inspect the gas circuit system for airtightness and connections to ensure stable and safe gas supply.

6.2 Periodic Maintenance

  • Replace Chromatographic Columns: Replace chromatographic columns based on their lifespan and analytical performance to ensure separation efficiency.
  • Calibrate Instrument: Regularly calibrate the instrument using standard samples to ensure the accuracy and reliability of analysis results.
  • Software Upgrades: Stay updated with software upgrades and update the Eager Xperience software promptly to obtain improved operational experiences and data analysis functions.

6.3 Troubleshooting

  • Instrument Fails to Start: Check power connections, gas supply, and instrument status indicators to排除 (exclude) power and gas supply issues.
  • Abnormal Analysis Results: Inspect sample preparation, injection operations, and parameter settings to排除 (exclude) operational errors and improper parameter configurations.
  • Gas Circuit System Failures: Check the airtightness, connections, and gas flow settings of the gas circuit system to排除 (exclude) gas circuit system failures.
  • Detector Signal Abnormalities: Inspect detector settings, baseline levels, and signal stability to排除 (exclude) detector failures.

VII. Safety Precautions

7.1 Operational Safety

  • Wear Protective Gear: Wear lab coats, gloves, and goggles during operations to ensure personal safety.
  • Follow Operating Procedures: Strictly adhere to the manual’s and software’s instructions to avoid safety accidents caused by违规 (non-compliance) operations.
  • Ensure Adequate Ventilation: Ensure the laboratory is well-ventilated to prevent the accumulation of harmful gases, which could harm operators.

7.2 Instrument Safety

  • Regular Instrument Inspections: Conduct regular comprehensive inspections of the instrument to ensure it is in good working condition.
  • Avoid Vibrations and Impacts: Prevent vibrations and impacts to the instrument during handling and use to avoid damaging internal components.
  • Prompt Maintenance and Repairs: Contact professional maintenance personnel promptly for maintenance and repairs upon discovering instrument faults or abnormalities to ensure normal instrument operation.

VIII. Conclusion

The Thermo Scientific Flash 2000 Organic Elemental Analyzer is a highly efficient and precise instrument for elemental analysis. Through this comprehensive user guide, users can better understand and operate the instrument, ensuring accurate and reliable experimental results. During actual operations, users should strictly adhere to operating procedures and safety precautions, regularly maintain and service the instrument to extend its lifespan and improve analytical efficiency. Additionally, users should stay updated with software upgrades and technical support services to obtain an improved operational experience and data analysis capabilities.

Posted on

A Systematic Diagnostic and Troubleshooting Guide for Abnormal Startup of Malvern Mastersizer Particle Size Analyzer

——Practical Analysis Based on HandleException / Default Policy Software Errors

Abstract
The Malvern Mastersizer series of laser particle size analyzers are widely used in laboratories and industrial quality inspection fields. However, abnormalities during software startup are not uncommon. This paper provides an in-depth analysis of the typical error message “An unexpected exception occurred while calling HandleException with policy ‘Default Policy'” that occurs during the startup process. It dissects the issue from the perspectives of the software framework, runtime library dependencies, instrument hardware communication, Aero dry dispersion module, and the Windows system level, offering a complete diagnostic logic, troubleshooting process, and solution ideas for third-party maintenance engineers and equipment managers.

I. Introduction: Why is Malvern Mastersizer prone to startup abnormalities?
The Mastersizer series (including models 2000, 3000, and 3000E) are high-precision particle size testing devices that involve multiple modules such as optical measurement modules, laser optical path systems, expansion units, high-speed data acquisition cards, communication links, and PC software environments. An abnormality in any of these modules can lead to software startup failure. In particular, the Mastersizer 3000 software adopts the Microsoft .NET + Enterprise Library exception management framework, resulting in a complex exception structure that is prone to “HandleException” and “Default Policy” related errors.

II. Reproducing the Fault Phenomenon: What does the error message indicate?
When users start the software, they may see a pop-up window labeled “Application Error” with the message “An unexpected exception occurred while calling HandleException with policy ‘Default Policy’. Please check the event log for details about the exception.” This indicates the following:

  • An exception has been captured internally by the software, such as module initialization failure, configuration file reading failure, or device non-response.
  • The “Default Policy” that captures the exception has itself encountered an error. The software uses the Microsoft Enterprise Library Exception Handling Block, and when the default policy fails to execute, the software cannot continue to start.
  • Such errors do not necessarily directly prove instrument damage; they are more likely to reflect issues such as driver abnormalities, missing software dependencies, or disconnected communication links.

III. Analysis of the Mastersizer Software Startup Process: Understanding the root causes of faults from the source

  1. Software loading of its dependent DLLs
    This includes the .NET Framework, VC++ Runtime, Malvern core module DLLs, and Enterprise Library configuration files, among others. If any DLLs are missing or corrupted, startup abnormalities will occur.
  2. Software reading of configuration files
    This involves instrument model information, recently used module configurations, communication ports, laser initialization parameters, and dispersion module configurations. Reading failures will trigger exceptions.
  3. Instrument communication initialization
    The communication link for the Mastersizer 3000 may be USB, fiber optic, or RS-232. If the software does not receive a response from the instrument during the initialization stage, an exception will be thrown, especially when there are abnormalities in the Aero dry dispersion module.
  4. Optical system initialization
    Failure to turn on the laser drive, non-response from the optical path unit, or no return from the ADC data acquisition card can also lead to software startup failure.
  5. Software UI loading
    This stage is unlikely to cause HandleException unless there is damage to system fonts or abnormalities in Windows graphical components.

IV. Typical root causes that may lead to HandleException (ranked by probability)

  1. Instrument communication failure (highest probability)
    Examples include loose or damaged USB cables, use of incompatible USB-HUBs, uninstalled or corrupted USB drivers, and Aero modules that are not powered on or have internal communication board failures.
  2. Corrupted or missing .NET Framework (very common)
    The software relies on .NET 3.5 and .NET 4.0/4.5. Windows updates, viruses, or incorrect software uninstallation can damage these components.
  3. Missing VC++ runtime libraries (often overlooked but very critical)
    Malvern uses a large number of C++ modules internally, and missing VC++ Runtime libraries will prevent the program from loading.
  4. Corrupted local configuration files of Malvern software
    Corruption or formatting errors in files such as software.config, exception.config, and user.config can prevent the Enterprise Library from reading them, triggering Default Policy errors.
  5. Windows permission issues
    Examples include the program being unable to write to ProgramData, the software not having administrator privileges, or company IT-installed antivirus systems blocking access to key files.
  6. Host and dispersion hardware issues
    These include damage to the Aero fan module, inability of the control board to power on, abnormal sensor output, or interrupted data links.

V. Complete on-site troubleshooting process (standard operating procedure for engineers)
Step 1: Confirm physical connections and power-on status
Check all USB/fiber optic communication cables, unplug and replug them, avoid using USB-HUBs, confirm that both the Mastersizer host and Aero are powered on, and observe whether the LED indicators are normal.
Step 2: Restart the device and computer
The recommended sequence is to close the software, turn off the instrument, restart the computer, turn on the instrument, and then open the software. This is the reset method recommended by Malvern.
Step 3: Check the Windows event log (critical)
Navigate to “Event Viewer → Windows Logs → Application” and search for relevant logs such as Malvern, Mastersizer, .NET Runtime, and Application Error to obtain detailed exception sources.
Step 4: Repair system runtime libraries
Install .NET Framework 3.5, .NET Framework 4.0/4.5, and VC++ 2005/2008/2010/2012/2013 runtime libraries. You can use the Microsoft .NET Repair Tool and the Visual C++ Redistributable Package collection to perform repairs.
Step 5: Reset or delete software configuration files (commonly effective)
Delete the configuration files in the C:\Users\username\AppData\Local\Malvern\ and C:\ProgramData\Malvern\ directories. The software will automatically regenerate them.
Step 6: Reinstall the software (ultimate solution)
This is suitable for situations such as software corruption, abnormal configuration files, missing DLLs, or interference from enterprise antivirus software. A complete reinstallation will almost restore normal operation.

VI. Special case: Abnormalities caused by the Aero dry dispersion module
In the Mastersizer + Aero dry dispersion module combination system, the Aero contains components such as a motor drive, differential pressure sensor, control CPU board, and speed feedback system. If the Aero’s internal hardware is damaged, error messages such as “Unexpected exception” and “Failed to initialize module: Aero” will appear during the software initialization stage. If you observe no indicator lights when the Aero is powered on, no startup action of the suction fan, abnormal fan current, or non-operation of the internal fan on-site, the problem may be concentrated on damage to the Aero control board or fan drive board.

VII. Best advice for engineers

  • Confirm communication lines and device power-on status: Re-plug the communication lines and avoid using USB-HUBs.
  • Restart the device and computer: Follow the correct restart sequence.
  • Check the event log: Obtain detailed exception information.
  • Repair the .NET Framework and VC++ Runtime: Ensure that software dependencies are complete.
  • Exclude equipment hardware abnormalities (especially Aero): Focus on the fan, control board, and power module.
  • Reinstall the software if necessary: Use this as the final solution.

VIII. Conclusion: The essence and solution direction of Mastersizer startup abnormalities
The error “An unexpected exception occurred while calling HandleException with policy ‘Default Policy'” analyzed in this paper is, from a software structure perspective, a secondary exception caused by the failure of the software’s exception handling mechanism. However, the root causes often lie in system runtime libraries, drivers, configuration files, communication links, or abnormal initialization of instrument modules (especially Aero). Through a systematic diagnostic process, almost 100% of the problems can be located.

IX. Appendix: On-site troubleshooting checklist for engineers (printable)
✔ Communication check

  • Loose USB/fiber optic cables
  • Whether the HUB has been removed
  • Whether the instrument is properly powered on
    ✔ Software environment
  • .NET Framework 3.5/4.x
  • Integrity of VC++ Runtime
  • Whether the software has been blocked by enterprise antivirus software
    ✔ Windows system
  • Permissions
  • Event Viewer
  • Whether there are conflicting drivers
    ✔ Instrument hardware
  • Aero fan
  • Control board
  • Internal sensors
  • Host power module
    ✔ Software repair
  • Delete configuration files
  • Reinstall the software

X. Overall Summary
By technically dissecting the startup process of the Malvern Mastersizer particle size analyzer and analyzing the root causes of HandleException / Default Policy errors, it can be concluded that such faults are the result of a comprehensive failure in the coordination of the software, system, drivers, and instrument initialization processes. As long as engineers master the troubleshooting logic proposed in this paper, they can quickly locate and accurately repair most on-site abnormalities.

Posted on

Comprehensive User Guide for the Tianrui X-Ray Fluorescence Spectrometer EDX1800

I. In-Depth Product Understanding

Core Features

  • High Efficiency and Stability: Equipped with a new-generation high-voltage power supply and X-ray tube with a power of up to 75W, enhancing testing efficiency and reliability.
  • Flexible Adaptability: Featuring a down-illumination design, it allows for the electric switching of various collimators and filters to accommodate different testing scenarios.
  • Precise Positioning: A fine manual moving platform and a high-resolution probe improve analytical accuracy.
  • Comprehensive Safety Protection: The X-ray tube is well-shielded, resulting in virtually zero radiation. It is equipped with self-locking and emergency lock mechanisms for all-around protection.

Key Testing Specifications

  • Element Range: From sulfur (S) to uranium (U).
  • Detection Limit: Reaching as low as 1 ppm, with a content range of 1 ppm to 99.9%.
  • Repeatability: Repeatability of 0.1% for multiple measurements and long-term operational stability of 0.1%.
  • Environmental Requirements: Temperature range of 15°C to 30°C and a power supply of 220V ± 5V.

Main Application Areas

  • ROHS Testing: Accurately detects hazardous elements in electronic and electrical products.
  • Precious Metal Testing: Quickly and accurately determines the content of precious metals and jewelry.
  • Coating Measurement: Measures the thickness of metal coatings and the content of electroplating solutions and coatings.
  • Geological and Mineral Analysis: Performs full-element analysis suitable for mineral exploration.

Unboxing Inspection Points

  • Check Items: Verify the presence of the instrument host, mounting plate, and accessory kit (including power cord, USB extension cable, etc.).
  • Inspect Appearance: Ensure there are no dents, scratches, and that all accessories are intact and undamaged.
  • Prompt Contact: Report any issues to the dealer or manufacturer immediately.

II. Instrument Installation and Debugging

Installation Environment Requirements

  • Complete Equipment: Equipped with heating and cooling air conditioners, computers, and printers.
  • Suitable Environment: Free from water sources, heat sources, strong electromagnetic interference, flammable materials, and excessive dust accumulation; avoid direct sunlight.
  • Reasonable Location: Keep away from extremely humid or low-temperature areas and places prone to vibrations. Maintain a distance of at least 30 cm from walls on all sides.

Installation Precautions

  • Avoid Flammable Materials: Do not install near alcohol or paint thinners.
  • Stable Installation: Place on a stable and sturdy tabletop or support.
  • Minimize Interference: Keep away from strong electromagnetic interference sources, handle with care, and ensure good ventilation.

Instrument Connection Steps

  • Power Connection: Connect the power cord between the instrument and the power strip.
  • Data Cable Connection: Connect the data cable between the instrument and the computer.
  • USB Extension Cable: Connect to the dedicated USB slot for the camera.

Debugging Process

  • Power Debugging: Turn on the main power, instrument host power, and computer power in sequence, and check the indicator light status.
  • Software Installation and Debugging: Install the RoHS software, copy the “Configure” and “Data” folders, and install Office software.
  • Instrument Initialization Debugging: Start the software, enter the password, place the silver calibration sheet, and perform initialization.

III. Complete Instrument Operation Process

Pre-Operation Preparation

  • Personnel Preparation: Operators must be trained and wear protective gear.
  • Hardware Inspection: Check that all connections are intact and the sample chamber is clean.
  • Software Inspection: Start the software and check the interface and functional modules.

Basic Instrument Operations

  • Power On: Turn on the main power, instrument host power, and computer power in sequence.
  • Sample Placement: Open the sample chamber, place the sample, and close the chamber.
  • Sample Removal: Open the sample chamber, remove the sample, and close the chamber.

Detailed Software Operations

  • Software Launch: Double-click the software icon or start it from the Start menu.
  • Interface Introduction: Menu bar, toolbar, status bar, program bar, report bar, and spectrum display area.
  • Parameter Settings: Configure measurement time, preheating, initialization, collimator, etc.
  • Sample Testing: Prepare, set the time, select the program, start testing, and view results.
  • Result Saving and Printing: Save spectra, import to Excel, and print reports.
  • Result Observation: Content display, custom standard setting, and virtual spectrum observation.

Instrument Calibration Operations

  • Pre-Calibration Preparation: Warm up the instrument, prepare calibration samples, and set calibration conditions.
  • Scan Standard Sample Spectra: Create a new working curve, initialize, and scan sample spectra.
  • Edit Working Curve: Set element boundaries, calculate intensities, edit intensity and content values, and observe linearity.
  • Re-test Standard Samples: Measure standard samples and adjust the curve.
  • Data Backup: Backup the “Configure” and “Data” folders.

Software Uninstallation Operations

  • Pre-Uninstallation Preparation: Backup data.
  • Uninstallation Steps: Uninstall through the Control Panel or Start menu.

IV. Instrument Maintenance and Care

Daily Maintenance

  • Designated User: Assign a specific person for use and storage.
  • Keep Clean: Regularly wipe the instrument surface and sample chamber.
  • Environmental Cleanliness: Maintain a clean, dry, and well-ventilated work environment.
  • Check Connections: Regularly inspect connection cables.

Regular Maintenance

  • Preheat Initialization: Preheat for 30 minutes and then initialize each time the instrument is turned on.
  • Parameter Testing: Regularly test and adjust instrument parameters.
  • Check Cooling: Ensure the fan is functioning properly and cooling vents are unobstructed.
  • Long-Term Storage: Cover with a dust cover and turn off the power when not in use for extended periods.
  • Protect Detector: Avoid touching or damaging the detector measurement window.

Special Situation Handling

  • Liquid Spillage: Immediately turn off the power and contact an authorized service center.
  • Collision Impact: Stop using the instrument and inspect it for damage.
  • Humid Environment: Take dehumidification measures.

V. Common Fault Analysis and Handling

Hardware Faults

  • High-Voltage Indicator Light Not On: Check the power switch and contact for replacement of high-voltage components.
  • Unable to Connect Normally: Check data cables and interfaces, and contact for repair.
  • Printer Connection Failure: Replace interfaces and data cables, and install drivers.

Software Faults

  • Unable to Start Normally: Check installation, system, and connections; reinstall the software.
  • Abnormal Test Results: Check sample placement, program selection, working curve, preheating initialization, and external environment.
  • Software Error or Freezing: Check computer configuration, reinstall the software, and standardize operations.

Other Faults

  • Abnormal Noise or Smoking: Immediately turn off the power and contact for repair.
  • Poor Repeatability of Test Results: Ensure sample uniformity, extend measurement time, stabilize preheating, recalibrate the curve, and clean the sample chamber.

VI. Safety Precautions

Installation Safety

  • Avoid Flammable Materials: Do not install near flammable items.
  • Stable Installation: Place on a stable and sturdy tabletop or support.
  • Suitable Environment: Avoid damp, dusty, sunny, high-temperature, or near open flame areas.

Operation Safety

  • Correct Power Plugging/Unplugging: Fully insert into sockets, keep away from heat sources, and hold the plug to unplug.
  • Prohibited Operations: Do not disassemble or modify the instrument, damage power cords, or use non-compliant voltages.
  • Voltage Stabilization: Use a voltage stabilizer to ensure stable voltage.
  • Abnormal Handling: Immediately turn off the power upon detecting abnormalities.
  • Protective Gear: Operators must wear protective gear; keep children and pregnant women away.

Environmental Safety

  • Compliance Requirements: Ensure the work environment meets temperature, humidity, air pressure, and power supply adaptability requirements.
  • Avoid Interference: Avoid strong electromagnetic interference during operation.
  • Good Ventilation: Maintain good ventilation in the work environment.

VII. Warranty Terms Explanation

  • Warranty Period: Free warranty for 12 months from the date of purchase.
  • Warranty Coverage: Only applies to the original consumer purchaser and is valid only in the country (or region) where the product was intentionally sold.
  • Warranty Service: Repair or replace defective products or parts free of charge; no charge for replaced parts, circuit boards, or equipment.
  • Post-Repair Warranty: Repaired products continue to enjoy warranty service for the remaining period of the original warranty.
  • Proof of Purchase: Consumers must provide purchase receipts or other proof.
  • Non-Warranty Situations: Non-normal use, improper storage, unauthorized modifications, etc.
  • Warranty Handling: Contact the purchase location or authorized service center; charges apply after the warranty period.

The Tianrui X-Ray Fluorescence Spectrometer EDX1800 is powerful and stable in performance. Users must strictly adhere to operational norms and maintenance requirements to ensure long-term stable operation of the instrument and obtain accurate and reliable test results. For difficult issues, it is recommended to consult the manual or contact an authorized service center for professional support.

Posted on

Complete User Guide for Beckman Cydem VT Automated Cell Culture System

Introduction

The Cydem VT Automated Cell Culture System, as a vital tool in modern biotechnology, significantly enhances the efficiency and stability of cell culture through its highly integrated automation design. Based on the core content of the system manual and combined with operational logic and practical tips, this guide provides researchers with a comprehensive reference for use. Covering the entire process from system overview to advanced applications, including installation, operation, maintenance, and troubleshooting, it aims to help users fully master the operation essence of this advanced equipment. The following content is strictly written in accordance with the manual specifications to ensure practicality and accuracy.

Chapter 1: System Overview and Core Advantages

1.1 System Definition and Application Scope

The Cydem VT system is a modular, fully automated cell culture platform that integrates four core modules: temperature control, gas regulation, liquid handling, and real-time monitoring. Designed to replace traditional manual operations, it is suitable for scenarios requiring high repeatability and sterile conditions, such as pharmaceutical research and development, oncology research, and stem cell culture. The system enables human-machine interaction through a touchscreen interface and remote control software, supporting multi-task parallel processing.

1.2 Technical Features Analysis

  • Precise Environmental Control: The incubator maintains a temperature fluctuation range of ≤ ±0.2°C, CO₂ concentration control accuracy of ±0.1%, and humidity above 95%, ensuring a stable environment for cell growth.
  • Automated Liquid Handling: Equipped with a built-in multi-channel pipetting arm, it supports liquid transfer from 1 μL to 50 mL with an error rate below 2%.
  • Contamination Prevention Mechanism: It employs a dual safeguard of HEPA filtration and UV sterilization, with key pipelines equipped with check valves to prevent cross-contamination.
  • Data Traceability Function: All operational parameters and cell images are automatically stored and can be exported in CSV or PDF formats.

Chapter 2: Hardware Installation and Initial Configuration

2.1 Site Preparation Requirements

The system should be placed on a level and stable laboratory bench with a surrounding clearance of at least 50 cm for heat dissipation. The power supply requirement is 220 V ± 10%/50 Hz, and an independent grounding line must be connected. The ambient temperature is recommended to be maintained between 18°C and 25°C, avoiding direct sunlight or direct alignment with ventilation openings.

2.2 Core Component Installation Process

  • Main Unit Positioning: Remove the transportation fixing bolts and adjust the feet until the level indicator shows green.
  • Culture Module Assembly: Insert the culture dish holder into the slide rail until it locks into place with a click. Handle glass components gently.
  • Liquid Pathway Connection:
    • Connect the culture medium bottle and waste liquid bottle to the color-coded interfaces respectively (blue for air intake, red for liquid pathway).
    • Perform pipeline priming: Select “Liquid Pathway Cleaning” in the software interface until there are no air bubbles in the pipeline.
  • Gas Source Configuration: Connect the CO₂ cylinder to the back interface of the system through a pressure reducer, with an initial pressure setting recommended at 0.1–0.15 MPa.

2.3 First-time Startup and Calibration

After powering on, the system performs a self-check (approximately 5 minutes), and the touchscreen displays the initialization interface. Follow the prompts to complete:

  • Sensor Calibration: Including pH electrode calibration (using standard buffer solutions) and O₂ probe calibration (zeroing in air).
  • Mechanical Arm Origin Correction: The pipetting arm automatically moves to the preset position and records the coordinates.
  • User Permission Settings: Assign administrator and operator accounts, set passwords, and define operational scope restrictions.

Chapter 3: Full Process Analysis of Daily Operations

3.1 Culture Initiation Phase

  • Step 1 – Program Creation: Create a new task in the “Protocol Editor,” with key parameters including:
    • Culture type (adherent/suspension cells)
    • Liquid exchange frequency (e.g., every 48 hours)
    • Termination conditions (OD value ≥ 0.8 or time threshold)
  • Step 2 – Sample Loading:
    • Use sterile forceps to place the culture dish on the loading platform and scan the barcode to associate sample information.
    • For adherent cells, allow them to settle for 10 minutes; for suspension cells, directly initiate the mixing program.
  • Step 3 – Environmental Parameter Setting: Select a preset mode according to the cell type (e.g., the HEK-293 mode automatically sets to 37°C/5% CO₂), or manually input:
Temperature: 37.0°C  
CO₂: 5.0%  
O₂: Set as required (conventionally 20%)  
Humidity: ≥ 95%

3.2 Monitoring During Operation

  • Real-time Data Viewing: Switch to the “Monitoring” tab on the main interface to view temperature fluctuation curves and pH trend graphs.
  • Abnormal Alarm Handling: When a “Liquid Insufficient” warning appears, pause the task → replace the culture medium bottle → resume operation.
  • Intermediate Intervention Operations: Wear sterile gloves, pause the mechanical arm using the emergency stop button, and quickly complete sampling or liquid supplementation.

3.3 Culture Termination and Sample Collection

Select the target experiment from the task list and click “Terminate.” The system automatically performs:

  • The pipetting arm aspirates and discards the waste liquid.
  • It injects 0.25% trypsin (for adherent cells).
  • The low-temperature preservation module is lowered to 4°C.
    After removing the samples, immediately execute the “Quick Clean” program (taking approximately 15 minutes).

Chapter 4: Maintenance and Upkeep Specifications

4.1 Daily Maintenance Checklist

  • Check the waste liquid bottle level (empty if it exceeds 80%).
  • Wipe the touchscreen and exterior surfaces with 70% ethanol.
  • Confirm the remaining pressure in the CO₂ cylinder (replace if it is below 0.05 MPa).

4.2 Weekly In-depth Maintenance

  • Pipeline Disinfection: Run the “Sterilization” program and circulate 0.1 M NaOH solution for 30 minutes.
  • Mechanical Arm Lubrication: Apply specialized silicone grease to the XYZ-axis guide rails (never use Vaseline).
  • Sensor Calibration: Soak the pH electrode in 3 M KCl storage solution and perform air calibration for the O₂ sensor.

4.3 Monthly Inspection Items

  • Replace the HEPA filter (Part Number: CYD-FIL-01).
  • Check the aging of the sealing rings of the pipette tips.
  • Back up system logs and user data to an external storage device.

Chapter 5: Fault Diagnosis and Emergency Response

5.1 Common Alarm Handling Solutions

Alarm CodeMeaningHandling Action
E-102Temperature Exceeding LimitCheck the incubator door seal and reset the heating module.
E-205Liquid Pathway BlockageExecute the pipeline backflush program and replace the 0.22 μm filter.
E-311Communication TimeoutRestart the control computer and check the network cable connection.

5.2 Emergency Situation Response

  • Power Interruption: The system automatically activates the backup battery to maintain the operation of key sensors. Power must be restored within 2 hours.
  • Contamination Incident: Immediately initiate “Emergency Sterilization” (UV + 75% ethanol spray). Contaminated culture dishes must be autoclaved before disposal.
  • Mechanical Arm Collision: Enter “Maintenance Mode” to manually adjust the arm position and calibrate the track encoder.

Chapter 6: Advanced Functions and Application Expansion

6.1 Multi-task Parallel Strategy

Through the “Batch Scheduler” function, up to 6 independent experiments can be managed simultaneously. It is recommended to group them according to the following principles:

  • Arrange the same type of cells in the same batch.
  • Prioritize high-frequency detection tasks for daytime periods.
  • Set resource conflict warnings (e.g., detection of overlapping pipette usage).

6.2 Data In-depth Analysis Techniques

  • Growth Curve Fitting: After exporting OD data, use the built-in Gompertz model in the system to calculate the doubling time.
  • Morphological Analysis: Combine with the microscopic imaging module to quantify cell aggregation degree through image segmentation algorithms.
  • Custom Report Template: In the “Report Generator,” drag and drop fields to generate experimental reports compliant with GLP specifications.

6.3 Remote Control Configuration

After connecting to the laboratory local area network via Ethernet:

  • Enable “Remote Access” permissions in the administrator account.
  • Use the official app (Cydem Controller) to scan the device QR code for binding.
  • Set operation delay compensation (recommended ≤ 200 ms within the local area network).

Conclusion

The value of the Cydem VT system lies not only in replacing manual operations with automation but also in ensuring the repeatability and traceability of experimental data through standardized processes. It is recommended that users establish a complete set of SOP documentation, participate in technical training organized by the manufacturer at least once a year, and stay updated on firmware update announcements to obtain functional optimizations. This guide covers the core operational scenarios of the system, and parameters should be flexibly adjusted according to specific experimental needs in actual use to maximize equipment performance.

Posted on

User Guide for Hitachi Ion Sputter Coater MC1000/MC100 Series

1. Introduction and Instrument Overview

The Hitachi MC1000 Ion Sputter Coater is a benchtop magnetron sputtering coating device specifically designed by Hitachi High-Tech Corporation for the preparation of scanning electron microscope (SEM) samples. It is used to deposit extremely thin (1 – 30 nm) conductive metal films on the surfaces of non-conductive samples, eliminating the charging effect during SEM observation and improving the quality of secondary electron imaging.

Core Advantages:

  • Utilizes magnetron sputtering technology to achieve low-temperature, low-damage, and high-particle-fineness coating.
  • Particularly friendly to heat-sensitive, biological, polymer, and other sensitive samples.
  • Features a 7-inch color LCD touch screen for operation and supports multiple languages.
  • The Recipe function allows for the storage of multiple sets of commonly used parameters for one-click recall.
  • Supports an optional film thickness monitoring unit for precise control of film thickness.
  • Highly modern and automated operation.
  • Applicable in fields such as materials science, biology, geology, semiconductors, nanotechnology, and failure analysis.
smart

2. Safety Precautions

Argon Gas Safety:

  • Ensure the operating environment is well-ventilated or install an oxygen concentration detector.

High-Voltage Electrical Risk:

  • Never open the cover or touch internal components during operation.

Vacuum Safety:

  • Always break the vacuum before opening the sample chamber.

Target Material Toxicity:

  • Wear gloves and a mask when replacing target materials.

Radiation:

  • A small amount of X-rays is generated during the sputtering process, but the equipment is shielded.

Prohibited Actions:

  • Never use oxygen or other active gases.
  • Do not place flammable, explosive, or strongly magnetic substances on the sample stage.
  • Do not leave the equipment unattended during operation.

Emergency Situations:

  • Immediately cut off the power supply, close the main argon gas valve, and evacuate personnel.

3. Technical Specifications

ItemSpecification Details
ModelMC1000
Sputtering MethodDC magnetron sputtering
Target Sizeφ50 mm × 0.5 mm
Sample StageStandard φ50 – 60 mm, rotatable; maximum sample height 20 mm
Target-Sample DistanceFixed at 30 mm
Ultimate Vacuum≤5×10⁻⁴ Pa
Working Vacuum5 – 10 Pa
Sputtering CurrentAdjustable from 0 – 40 mA
Sputtering VoltageAdjustable from 0 – 1.5 kV
Coating RateAu: ~35 nm/min; Au/Pd: ~25 nm/min; Pt: ~15 nm/min; Pt/Pd: ~20 nm/min
Film Thickness ControlTime control or optional film thickness meter
Vacuum PumpTurbo molecular pump + rotary mechanical pump
Operating GasHigh-purity argon (above 99.99%)
Gas Flow ControlAutomatic mass flow controller (MFC)
Display/Operation7-inch color LCD touch screen
Recipe StorageUp to 5 – 10 sets
Power SupplyAC 100 – 240 V, 50/60 Hz, single-phase, approximately 1.5 kVA
DimensionsApproximately 450 (W) × 391 (D) × 390 (H) mm
WeightMain unit approximately 25 kg, pump set approximately 28 kg
Operating EnvironmentTemperature 15 – 30℃, humidity ≤85% (no condensation)

4. Instrument Structure and Panel Description

Front View:

  • 7-inch touch screen
  • Sample chamber glass cylinder
  • Target height adjustment knob (present on some older models)
  • Main power switch

Rear Panel:

  • Argon gas inlet
  • Vacuum pump power and signal lines
  • Main power socket
  • Exhaust port

Internal Structure:

  • Magnetron target
  • Sample stage
  • Quartz crystal oscillator film thickness probe (optional)

5. Installation and First-Time Startup Preparation

  • Place the equipment on a stable laboratory bench, away from vibration sources.
  • Use a three-prong socket with a ground wire, with a grounding resistance ≤100 Ω.
  • Connect the argon gas cylinder and set the secondary pressure to 0.03 – 0.05 MPa.
  • Check the vacuum pump oil level.
  • Conduct an initial vacuum pumping test and observe whether it reaches the 10⁻³ Pa level.

6. Detailed Operation Steps

6.1 Startup and Preparation

  • Open the main valve of the argon gas cylinder and set the secondary pressure to 0.04 MPa.
  • Connect the main unit power.
  • The touch screen lights up, and the main interface is displayed.

6.2 Sample Placement

  • Ensure the chamber is vented to atmospheric pressure.
  • Lift the glass cylinder cover.
  • Secure the sample on the sample stage.
  • Adjust the target-sample distance.
  • Close the glass cylinder.

6.3 Parameter Setting

  • Click “Process” or “Recipe”.
  • Set parameters such as target material type, sputtering current, and sputtering time or film thickness.
  • Save as a Recipe.

6.4 Starting Coating

  • Click “START”.
  • The equipment automatically performs the coating process.

6.5 Sample Retrieval and Shutdown

  • After coating is complete, the equipment automatically breaks the vacuum.
  • Open the glass cylinder and remove the sample.
  • Close the glass cylinder, click “Vent” or long-press “STOP”.
  • Turn off the power switch and close the main valve of the argon gas cylinder.

7. Recommended Common Recipe Parameters

Application ScenarioTarget MaterialCurrent (mA)Time (s)Estimated Film Thickness (nm)Remarks
Conventional SEMAu20608 – 12Economical
High-resolution FE-SEMPt or Pt/Pd25905 – 10Finest particles
Biological SamplesAu/Pd15 – 2012010 – 15Low-temperature priority
EDS Energy-Dispersive Spectroscopy AnalysisCarbon evaporation (optional)10 – 20Avoid metal peak interference
Thick or Large SamplesAu3018020 – 30Requires optional large chamber

8. Target Replacement Steps

  • Completely break the vacuum and open the glass cylinder.
  • Wear gloves and use an Allen wrench to loosen the target pressure ring.
  • Remove the old target material.
  • Place the new target material.
  • Tighten the pressure ring.
  • Close the glass cylinder, pump down the vacuum, and check for leaks.
  • Run an empty coating process once.
  • Target Lifespan: An Au target can typically be used for approximately 500 – 800 coating sessions.

9. Daily Maintenance and Care

Maintenance ItemFrequencyMethod
Cleaning the Sample Chamber Glass CylinderAfter each useWipe with a lint-free cloth and isopropyl alcohol or acetone
Checking O-ringsWeeklyVisually inspect and lightly coat with silicone grease
Replacing Vacuum Pump OilEvery 300 – 500 hoursDrain the oil → Clean the oil tank → Add new oil
Molecular Pump MaintenanceEvery 1 – 2 yearsReturn to the factory or have a professional regenerate it
Cleaning the Target SurfaceWhen replacing the targetPolish the oxide layer with fine sandpaper
Overall Dust RemovalMonthlyClean with a vacuum cleaner and a soft brush
Checking the Argon Gas PipelineMonthlyCheck for leaks at the joints

10. Common Troubleshooting

Fault PhenomenonPossible CausesSolutions
Failure to IgniteInsufficient argon pressure / Target oxidation / Excessive vacuumCheck the argon pressure; perform an empty coating to remove oxidation; reduce the vacuum
Unstable or Low CurrentDepleted target material / Poor contactReplace the target material; check the tightness of the pressure ring
Inability to Achieve VacuumInsufficient pump oil / Leakage / Aging O-ringsAdd pump oil; check for leaks; replace O-rings
Discrepancy Between Coated Film Thickness and Set ValueDirty film thickness meter probe / Change in target material coating rateClean the quartz crystal oscillator; recalibrate the film thickness meter
Unresponsive Touch ScreenPower fluctuations / Software crashRestart the main unit; contact after-sales service
Sample Overheating or DamageExcessive current / Prolonged coating timeReduce the current; perform coating in multiple sessions

11. Optional Accessories Introduction

  • Film Thickness Monitoring Unit: Real-time measurement using a quartz crystal oscillator with an accuracy of ±0.1 nm.
  • Large Sample Chamber: Sample diameter up to 150 mm and height 30 – 50 mm.
  • Carbon Evaporation Attachment: Used for EDS analysis.
  • Various Target Materials: Pt, Au/Pd, Pt/Pd, etc.
  • Automatic Transformer: Supports a wide voltage range of 115 – 240 V.

12. Precautions and Best Practices

  • A new target material must undergo an empty coating process for 20 – 30 seconds during its first use.
  • For biological samples, it is recommended to use a Pt target with a low current.
  • When the equipment is not in use for an extended period, start it up and pump down the vacuum for 1 hour every week.
  • Record the coating parameters and SEM imaging results for each session.
  • For the complete official Chinese manual, please contact Hitachi High-Tech China or local agents.
Posted on

Yokogawa Recorder SR10000 User Guide

Foreword

The SR10000 series recorders produced by Yokogawa Electric Corporation are high-performance, multi-channel data recording devices widely used in industrial process control, laboratory monitoring, and other fields. This guide aims to systematically organize the official manual, extract key operations, and help users quickly master and effectively apply the recorder.

Chapter 1: Device Overview and Core Concepts

1.1 Models and Basic Parameters

  • Model Classification: Pen-type (SR10001 – SR10004) and dot-matrix type (SR10006).
  • Measurement Cycle: The pen-type has a fixed measurement cycle of 125 ms, while the dot-matrix type depends on the A/D integration time.
  • Input Channels: Correspond to the number of pens or dots in the model. Unused channels can be set to “Skip”.

1.2 Two Operating Modes

  • Setting Mode: Press and hold the MENU key for 3 seconds to enter and set daily parameters.
  • Basic Setting Mode: In the setting mode, press and hold the △ and ▽ keys simultaneously for 3 seconds to enter for in-depth system configuration.
  • Important Note: The basic setting mode cannot be accessed during recording.

1.3 Core Concepts

  • Range Type: Such as thermocouple type K, DC voltage 2V, etc., with fixed measurable ranges.
  • Input Range: Specify the actual measurement range within the measurable range.
  • Recording Range: On the recording paper, a width of 100 mm represents 0% to 100% of the input range.
  • Scale Calculation: Linearly convert voltage signals into actual physical units.

Chapter 2: Detailed Explanation and Configuration of Measurement Input Functions

2.1 Input Type and Range Setting

  • Operation Path: Setting mode → RANGE, select the channel and input type, and set the range values.

2.2 Input Signal Processing Functions

  • Filter (Pen-type Models): A low-pass filter to smooth signals.
  • Moving Average (Dot-matrix Models): Calculate the average of consecutive sampled values.
  • A/D Converter Integration Time: Suppress power frequency interference.

2.3 Advanced Calculation and Compensation Functions

  • Bias: Add a fixed offset to the measured value.
  • Input Value Calibration (/CC1 Optional Accessory): Multi-point broken-line calibration.
  • Thermocouple Cold Junction Compensation: Compensate for errors caused by cold junction temperature changes.
  • Thermocouple/1 – 5V Open-circuit Detection: Detect signal disconnections and trigger alarms.

Chapter 3: Alarm Function Configuration and Management

3.1 Alarm Types and Setting

  • Operation Path: Setting mode → ALARM, select the channel and alarm number, and set the alarm type and value.

3.2 Advanced Alarm Settings

  • Alarm Hysteresis: Prevent frequent alarm operations.
  • Alarm Output Relay Action: Select the action mode of the relay when an alarm occurs.
  • Diagnostic Output: Trigger relay 101 when the recorder fails.

Chapter 4: Comprehensive Analysis of Recording and Printing Functions

4.1 Curve Recording

  • Pen-type Models: Continuous recording with fixed colors.
  • Dot-matrix Models: Periodic dot-matrix recording, with adjustable recording cycles and colors.

4.2 Paper Feed Speed and Area Recording

  • Paper Feed Speed: Setting mode → CHART, select the speed gear.
  • Area Recording: Limit the recording range for specific channels.

4.3 Printing Output Functions

  • Timed Printing: Print at set time intervals.
  • Alarm Printing: Print when an alarm occurs or is cleared.
  • Information Printing: Print preset information triggered by manual or remote signals.
  • Manual Printing, List Printing, and Setting List Printing: Meet different printing needs.

4.4 Advanced Recording Functions

  • Partial Compression/Expansion Recording: Compress or expand a specific part of the recording range for display.
  • Phase-synchronized Recording (Pen-type Models): Ensure that the recording times of multiple pens are aligned.

Chapter 5: Maintenance, Calibration, and Troubleshooting

5.1 Regular Inspection and Cleaning

  • Inspection Items: Display, recording, and printing functions, and the remaining amount of recording paper.
  • Cleaning: Regularly clean the transmission shaft of the writing pen holder or the printing pen holder.

5.2 Calibration

  • Calibration Instruments: High-precision standard signal generators, etc.
  • Calibration Steps: Connect the device, warm it up, input standard signals, and check the displayed and recorded values.

5.3 Pen Position/Dot Position Adjustment

  • Pen-type Models: Basic setting mode → P_ADJ, adjust the left and right end positions.
  • Dot-matrix Models: Adjust the central hysteresis, left end, and right end positions in sequence.

5.4 Troubleshooting

  • Check Error Messages: The display shows error codes, which can be solved by referring to the manual.
  • Use Troubleshooting Flowcharts: Diagnose common problems.
  • Common Problem Checkpoints: Power supply, input signals, recording paper/pens/ribbons, key locks/custom menus.

Chapter 6: Introduction to Optional Accessories and Advanced Functions

  • Communication Function (/C3, /C7): Remote monitoring and data acquisition.
  • Alarm Output Relays (/A1, /A2, /A3): Provide more relay output points.
  • Remote Control (/R1): Control recorder functions with external signals.
  • Title Printing (/BT1): Enhance batch printing functions.
  • Input Value Calibration (/CC1): High-precision multi-point broken-line calibration.
  • Extended Input (/N1, /N3): Support more thermocouple and thermistor types.

Chapter 7: Summary and Best Practice Recommendations

  • Plan Before Setting: Clearly define the setting requirements for each channel.
  • Make Good Use of the Function Setting Wizard: It provides great help for complex configurations.
  • Pay Attention to Signal Quality: Ensure correct wiring, grounding, and filtering.
  • Use Printing Functions Reasonably: Set printing intervals and event markers according to needs.
  • Establish a Maintenance Calendar: Regularly clean, inspect, and calibrate.
  • Operate Safely: Perform wiring and maintenance after power-off and comply with safety signs.
Posted on

From Waterway Blockage to Light Path Deviation: A Comprehensive Analysis of System Diagnosis and Maintenance Process for Abnormal Performance of Anton Paar PSA 1090 LD Laser Particle Size Analyzer

Abstract

Laser particle size analyzers are widely used in fields such as materials science, powder technology, biopharmaceuticals, and mineral processing. Their measurement accuracy and repeatability are key indicators for evaluating equipment performance. The Anton Paar PSA 1090 LD, as a high-precision wet laser particle size analyzer, may encounter typical abnormalities such as “slow drainage, low flow rate, system blockage, poor measurement repeatability, and large particle size deviation” during long-term use. Based on actual fault cases of a user’s equipment, this study conducts a systematic analysis from multiple dimensions including the light path, flow path, circulation pump, dispersion cell, and drainage channel, and proposes technical cause determination methods and engineering maintenance steps. This article aims to provide a complete set of fault diagnosis methods and scientific maintenance paths for third-party laboratories, after-sales engineers, and equipment users, helping to improve instrument reliability and service life.

1. Introduction

Laser particle size analyzers play an irreplaceable role in the field of powder and particle material characterization. With the rapid development of materials science and nanotechnology, the requirements for the accuracy, stability, and repeatability of particle size testing continue to increase. The Anton Paar PSA 1090 LD, as an internationally recognized laser particle size analyzer, has core advantages such as high light path stability, good dispersion effect, and high system automation. However, even high-end equipment may still encounter typical problems such as “slow drainage, blockage, poor repeatability, and large particle size deviation” during long-term operation or improper maintenance.

Based on real-world usage cases, this article, from the perspective of third-party laboratory engineers, systematically analyzes the root causes of such faults and provides immediately implementable diagnostic methods, aiming to provide high-value references for relevant practitioners.

2. Working Principle and System Composition of the PSA 1090 LD

To understand why the equipment exhibits abnormalities, it is necessary to first understand its internal structure and operating mechanism.

2.1 Introduction to the Wet Dispersion System

The PSA 1090 LD uses a wet dispersion method, where the liquid is driven by a circulation pump to form a continuous flow between the sample cell and the water tank. The water flow undertakes three tasks:

  • Transporting sample particles
  • Ensuring uniform dispersion of particles
  • Providing a stable light path environment

The stability of the flow rate determines whether the sample can uniformly pass through the light beam and whether the measurement can be precise.

2.2 Structure of the Light Path System

The laser is emitted from the transmitting end, passes through the sample in the sample cell, and the scattered light is collected by the detector. If the light path is affected, it will lead to significant data deviations.

Light path window contamination may cause:

  • Unstable scattered light intensity
  • Increased data noise
  • Abnormal oscillation of the particle size curve

This is an important factor contributing to measurement deviations.

2.3 Importance of the Circulation System and Fluid Dynamics

The circulation system consists of:

  • Suction hose
  • Circulation pump
  • Flow cell (sample cell)
  • Drainage channel

An increase in resistance at any position will lead to:

  • Decreased water flow
  • Inability to discharge bubbles
  • Accumulation of particles in the cell
  • Unstable test curves

Actual cases show that fluid dynamic problems are the main source of abnormalities in the PSA series.

3. Fault Manifestations and Initial Symptoms

According to feedback from the user’s site and video footage, the equipment exhibited typical system fault characteristics.

3.1 Slow Drainage and Insufficient Flow Rate

This is the most intuitive abnormal phenomenon. A normal device should be able to complete drainage quickly, but in this case:

  • The drainage speed is significantly reduced
  • The water flow is interrupted or intermittent
  • There is a noticeable sense of resistance

This indicates partial blockage within the circulation system.

3.2 Particle Deposition and Flocculation in the Sample Cell

From the photos of the sample cell window, it can be seen that:

  • There is a large amount of sediment at the bottom
  • There are flocculent impurities
  • The light path channel is not clean

This directly affects measurement accuracy.

3.3 Huge Deviations in Multiple Measurement Results

For example:

  • D50 changes from 0.8 µm to 58 µm (a jump of 70 times)
  • The shapes of the three curves are completely different

This phenomenon is definitely not due to sample problems but rather:

  • Uneven flow rate
  • Incomplete dispersion of aggregates
  • Laser signal fluctuations

These cause systematic deviations.

3.4 Bubble Retention and Discontinuous Fluid Flow

The video shows the presence of:

  • A large number of bubbles in the liquid
  • Interruptions and jumps in the liquid flow
  • Inability of the water body to continuously flow through the sample cell

This directly leads to a sharp increase in optical signal noise.

4. Systematic Analysis of Fault Causes

Based on the fault manifestations, the main abnormal sources involved in this case are as follows.

4.1 Blockage in the Dispersion Cell and Flow Cell

The bottom of the sample cell and the drainage outlet are the most prone to blockage. Long-term accumulation of:

  • Microparticles
  • Scale
  • Sediment
  • Organic film

will narrow the fluid channel.

Results:

  • Insufficient flow rate
  • Discontinuous signals
  • Jittering of the particle size curve

4.2 Blockage in the Drainage Channel (Core Cause in This Case)

The drainage channel is narrow, and even a small amount of sediment can significantly affect the flow rate. In this case, the obvious slowdown in drainage indicates severe blockage in the channel.

4.3 Insufficient Suction or Excessive Load of the Circulation Pump

The circulation pump is not damaged but rather:

  • The resistance in the pathway has increased
  • It is difficult to form sufficient flow
  • The pump idles, is sluggish, or has fluctuating water output

This leads to abnormalities in the entire system.

4.4 Aging of the Water Inlet Hose and Formation of Biofilm

The hose in this case has shown:

  • Yellowing
  • Rough inner walls
  • Increased flow resistance

Biofilm or sediment reduces the water absorption efficiency.

4.5 Light Path Window Contamination and Optical Signal Attenuation

Deposits on the window will:

  • Change the incident light intensity
  • Cause abnormal scattering
  • Trigger abnormal peaks in particle size
  • Deform the distribution curve

This is significantly present in this case.

4.6 Software Parameter Factors

Although parameters such as refractive index and dispersion mode can also affect the results, they will not cause mechanical problems such as “slow drainage” and can be excluded.

5. Engineering Diagnostic Steps

The following diagnostic process can be used by third-party laboratories to judge the performance of the PSA series wet systems.

5.1 Flow Observation Method

Normal: Continuous flow
Abnormal: Flow interruption, slowness, repeated appearance of bubbles
In this case, the flow rate is severely insufficient.

5.2 Blank Baseline Stability Judgment

A stable signal during blank testing indicates a normal light path; fluctuations suggest light path or fluid abnormalities.
In this case, the baseline noise is significantly increased.

5.3 Evaluation of Ultrasonic Dispersion Effectiveness

If particles still aggregate after ultrasonic activation, it indicates:

  • Insufficient flow rate
  • Inability to carry away aggregates

rather than a fault in the ultrasonic device itself.

5.4 Inspection of the Optical Window of the Sample Cell

The presence of:

  • Mildew spots
  • Scale
  • Contamination points

may lead to unstable data.

5.5 Drainage Speed Test

The slower the drainage speed, the more it indicates:

  • Blockage in the flow channel
  • Adherents on the pipe walls
  • Excessive system resistance

In this case, the drainage speed has significantly decreased.

5.6 Judgment of Circulation Pump Performance

If the pump can operate normally but the flow rate is insufficient, it is mostly due to excessive resistance, and the pump may not necessarily be damaged.

6. System Maintenance and Recovery Plan (Engineer Level)

The following are the most effective maintenance steps for the PSA series.

6.1 Cleaning the Flow Path: Circulation with 1% NaOH Solution

Steps:

  • Add 1% NaOH solution to the water tank
  • Operate at the maximum flow rate for 10–15 minutes
  • Then rinse with a large amount of pure water for 10 minutes
  • If there is an ultrasonic function, activate it for collaborative cleaning

Functions:

  • Dissolve sediment
  • Remove biofilm
  • Clean the flow channel

6.2 Reverse Flushing of the Sample Cell (Key Step)

Using a 50–100 mL syringe:

  • Unplug the drainage hose
  • Aim the syringe at the drainage outlet
  • Inject water backward into the sample cell

It is normal to flush out black or yellow sediment. This is the most effective unclogging method for the PSA series.

6.3 Replacement of the Water Inlet Hose and Drainage Pipe

Aging hoses cause poor water absorption. In this case, the pipes are obviously aged and need to be completely replaced with new ones.

6.4 Cleaning Method for the Light Path Window

Use:

  • 70–99% IPA
  • Fiber-free cotton swabs

Gently wipe the contaminated areas and avoid scratching with hard objects.

6.5 Standard Process for Eliminating Bubbles

  • Operate at the maximum circulation
  • Tilt the instrument by 20–30 degrees
  • Discharge the liquid multiple times
  • Continuously observe the changes in bubbles inside the sample cell

6.6 Final Calibration and Repeatability Verification

Test:

  • Three repeatability curves
  • Stability of D10, D50, and D90
  • Baseline noise level

After recovery, the curves should have a high degree of overlap.

7. Case Study: Correspondence between Abnormal Data and Real Causes

In this case, typical “data distortion caused by unstable system flow rate” is observed.

7.1 Abnormal Shoulder Peaks in the Particle Size Distribution Curve

Shoulder peaks indicate that the particles are not uniformly dispersed, which is a false peak caused by unstable flow.

7.2 Direct Correlation between D50 Jumps and Flow Rate Problems

Insufficient flow rate will lead to:

  • Deposition of large particles, resulting in false large particle peaks
  • Uneven concentration, causing jumps

This is completely consistent with this case.

7.3 Reasons for Different Shapes of Three Measurement Curves

  • Interruption of water flow
  • Bubbles passing through the light path
  • Fluctuations in sample concentration

Not due to the sample itself.

8. Preventive Maintenance Strategies and Recommendations

To prevent similar faults from occurring again, the following maintenance system should be established:

8.1 Lifespan Management of Pipelines

It is recommended to replace hoses every 6–12 months.

8.2 Flow Path Cleaning Plan

Recommendations:

  • Clean with pure water once a week
  • Perform NaOH circulation once a month
  • Conduct reverse flushing once a quarter

8.3 Light Path Maintenance Cycle

Check the light path window every 1–2 months and immediately remove any scale if present.

8.4 Water Quality and Environment

Must use:

  • Deionized water (electrical conductivity < 10 μS/cm)
  • Clean sample cups
  • Avoid dust entering the water tank

9. Conclusion

This case fully demonstrates that when the Anton Paar PSA 1090 LD exhibits faults such as “slow drainage, blockage, and large particle size deviation,” the root causes are mostly a combination of fluid dynamic abnormalities, light path contamination, and aging pipelines. Through systematic diagnosis and engineering maintenance, the equipment performance can be fully restored.

Key insights include:

  • The flow rate is the primary factor affecting the measurement accuracy of wet methods
  • The drainage channel and sample cell are the most important cleaning points
  • Light path window contamination can sharply reduce measurement repeatability
  • Pipeline aging can lead to potential resistance problems
  • Ultrasonication and flow rate must work in tandem to ensure sufficient dispersion

For third-party laboratories and engineers, establishing standardized maintenance procedures is a necessary measure to ensure the long-term stable operation of instruments.