I. Spectrometer Fault Handling
- Invalid Detection, No Data Output:
- Replace or upgrade the light source equipment.
- Repair or clean the detector.
- Realign the optical path.
- Fault Causes: Insufficient light emission from the source, damaged or dusty detector, misaligned or shifted optical path.
- Optical Path Obstruction:
- Clear the sample or optical path.
- Check for damaged components or dust in the optical path and clean or replace as necessary.
- Fault Causes: Sample residue in the optical path, blockages such as damaged components or dust.
- Vacuum Pump Not Starting:
- Ensure consistency between pump and oil models, and regularly replace pump oil.
- Adjust indoor temperature to avoid excessively low temperatures.
- Check and manually start the pump motor via the analysis software.
- Fault Causes: Poor viscosity or incorrect model of pump oil, temperature-related issues, control signal problems.
- Sudden Loss of Negative High Voltage:
- Try unplugging and replugging the second ground wire of the control box; if unsuccessful, consider restarting the instrument.
- Fault Cause: Vacuum value not rising for an extended period.
- Low Light Intensity:
- Clean the condensing lens.
- Clean the slit.
- Replace the corresponding photomultiplier tube.
- Fault Causes: Dirty condensing lens, dirty slit, degraded performance of the photomultiplier tube.
- Unstable Instrument Analysis Data:
- Check and replace the rear fan of the instrument if necessary.
- Fault Causes: Poor temperature control in the instrument’s vacuum chamber, faulty rear fan operation.
- Poor Communication Line Contact Leading to Computer Crash or Program Error:
- Reconnect the communication line.
- Exhaust Blockage:
- Replace the exhaust pipe with a transparent plastic tube and regularly purge the exhaust line.
- Fault Causes: Blocked argon exhaust line, debris in the lower bend of the spark chamber, debris at the inlet of the argon filter.
- Power, Connection, or Software Issues:
- Check for proper power connection, damaged or loose power cords.
- Ensure correct and undamaged connection of cables.
- If software issues occur, attempt reinstallation or updating.
- Other Maintenance and Calibration:
- Regularly clean the optical components of the spectrometer and perform calibration to ensure accuracy.

Note: The above solutions are for reference only. For specific faults, it is recommended to contact Rongji Electromechanical for professional repair and support.

II. Supported Spectrometer Brands by Rongji Electromechanical
- Niton Spectrometers (Thermo Fisher Scientific):
- Niton XL2, Niton XL2 Plus, Niton XL5 Plus, Niton DXL Precious Metal Analyzer, Niton XL2 100G General Metal Analyzer, Niton Apollo Handheld LIBS Analyzer.
- Yokogawa Spectrometers:
- AQ6370, AQ6370B, AQ670C, AQ6370D, AQ6374, AQ6375, AQ6376, AQ6315E, AQ6317B, AQ6317C, AQ6319, AQ6330.
- Hitachi Spectrometers:
- OE720, OE750, PMI-MASTER Smart, X-Supreme8000, LAB-X5000.
- HORIBA Spectrometers:
- FluoroMax Plus, VS70-PDA-HDR, Duetta, Ultima Expert, MicroHR Series, MicOS, 1000M Series, HR Evolution, iHR320, XploRA PLUS XS, T64000, HE.
- Malvern Spectrometers:
- Mastersizer 3000+ Ultra, Mastersizer 3000+ Pro, Mastersizer 3000+ Lab, Mastersizer 3000E.
- Beckman Coulter Spectrometers:
- DU 640, DU 720, DU 730.
- Shimadzu Spectrometers:
- ICPE-9000, ICPE-9800, ICPS-7000, ICPS-7510, ICPS-8100, EDX-LE Plus, EDX-GP, EDX-7200, EDX-8000, EDX-8100, IRSpirit, IRPrestige-21, PDA-5000, PDA-7000, PDA-8000, SolidSpec-3700, 3700DUV, RM-3000, RF-6000, UV-1900I.
- Thermo Fisher Scientific Spectrometers:
- Nicolet Summit, ARL 3460, ARL 4460, ARLeasySpark, DXRxi, DXR, DXR™ 3, WDXRF, EDXRF, XRFXRD, ARL EQUINOX.
- Bruker Spectrometers:
- VERTEX Series, ALPHA II, INVENIO, IFS 125HR, MPA II, TANGO, MATRIX-F, SENTERRA II, BRAVO, MultiRAM, RAM II, TXRF S4 TSTAR, S2 PICOFOX, M4 TORNADO, S1 TITAN, Q4 TASMAN, Q2, Q2L, Q8.
- Agilent Spectrometers:
- Agilent 8453, Agilent Cary 60, Agilent Cary 100, Agilent Cary 300, Agilent 55, Agilent 4100 MP-AES, Agilent 710/715 Series, Agilent 5110.
- HORIBA Additional Models:
- HR Evolution, iHR320, XploRA PLUS XS, T64000, HE.
- PerkinElmer Spectrometers:
- Spectrum Two, Spectrum 3, PinAAcle 500, PinAAcle 900F.
- Tianrui Instruments:
- EDX Series, SUPERXRF1050, Explorer8000 Series, SEE 100, SEE 200, ZSX Primus Series, XEPOS.
- Nanjing Sanxiang Spectrometers:
- THICK800A, EDX4500H.

Longi Electromechanical offers long-term maintenance for various types of spectrometers, including infrared, Raman, fluorescence, direct-reading, UV-Vis, and more. With nearly 30 years of experience, we can quickly repair various instruments. Additionally, we recycle and sell used spectrometers. For more information, please contact us.